Reliability and validity of openpose for measuring hip-knee-ankle angle in patients with knee osteoarthritis

Nature

Reliability and validity of openpose for measuring hip-knee-ankle angle in patients with knee osteoarthritis"


Play all audios:

Loading...

ABSTRACT We aimed to assess the reliability and validity of OpenPose, a posture estimation algorithm, for measuring hip-knee-ankle (HKA) angle in patients with knee osteoarthritis, by


comparing it with radiography. In this prospective study, we analysed 60 knees (30 patients) with knee osteoarthritis. We measured HKA angle using OpenPose and radiography before or after


total knee arthroplasty and assessed the test–retest reliability of each method with intraclass correlation coefficient (1, 1). We evaluated the ability to estimate the radiographic


measurement values from the OpenPose values using linear regression analysis and used intraclass correlation coefficients (2, 1) and Bland–Altman analyses to evaluate the agreement and error


between OpenPose and radiographic measurements. OpenPose had excellent test–retest reliability (intraclass correlation coefficient (1, 1) = 1.000) and excellent agreement with radiography


(intraclass correlation coefficient (2, 1) = 0.915), with regression analysis indicating a large correlation (R2 = 0.865). OpenPose also had a 1.1° fixed error and no systematic error when


compared with radiography. This is the first study to validate the use of OpenPose for the estimation of HKA angle in patients with knee osteoarthritis. OpenPose is a reliable and valid tool


for measuring HKA angle in patients with knee osteoarthritis. OpenPose, which enables non-invasive and simple measurements, may be a useful tool to assess changes in HKA angle and monitor


the progression and post-operative course of knee osteoarthritis. Furthermore, this validated tool can be used not only in clinics and hospitals, but also at home and in training gyms; thus,


its use could potentially be expanded to include self-assessment/monitoring. SIMILAR CONTENT BEING VIEWED BY OTHERS RELIABILITY OF HIP JOINT POSITION SENSE TESTS USING A CLINICALLY


APPLICABLE MEASUREMENT TOOL IN ELDERLY PARTICIPANTS WITH UNILATERAL HIP OSTEOARTHRITIS Article Open access 10 January 2022 RELIABILITY OF THE PELVIS AND FEMUR ANATOMICAL LANDMARKS AND


GEOMETRY WITH THE EOS SYSTEM BEFORE AND AFTER TOTAL HIP ARTHROPLASTY Article Open access 11 December 2022 ANALYSIS OF THE CONCURRENT VALIDITY AND RELIABILITY OF FIVE COMMON CLINICAL


GONIOMETRIC DEVICES Article Open access 27 November 2023 INTRODUCTION Hip-knee-ankle (HKA) angle is widely used as a measurement of knee alignment in knee osteoarthritis (OA). The HKA angle


is defined as the angle between the mechanical axes of the femur and tibia in the coronal plane1. Based on this angle, lower extremity alignment is classified as varus, neutral or


valgus2,3,4. Many reports have shown that varus or valgus alignment increases the risk of incidence and progression of knee OA5,6,7,8,9. Moreover, surgical treatment of knee OA, such as


osteotomy and arthroplasty, is important to reconstruct optimal alignment and correct load distribution in the medial and lateral compartments. In total knee arthroplasty (TKA), excessive


postoperative varus alignment has been reported to have a negative impact on patient satisfaction and implant longevity10,11,12,13,14. Therefore, the measurement of the HKA angle is


important throughout the entire course of knee OA. The HKA angle is measured from a full-length lower-limb radiograph in a standing position. Full-length radiographs involve the risk of


radiation exposure, additional technician training, increased time for image acquisition, and increased overall cost. Therefore, methods to estimate the HKA angle without full-length


radiographs have been discussed. One such method is to estimate the HKA angle from a standard anteroposterior (AP) knee radiograph15,16. However, the difference between the anatomical and


mechanical axes of the femur and tibia varies across patients, and this difference may reduce the accuracy of the estimation. In addition, since radiation exposure is unavoidable, routine


measurement of the HKA angle is difficult. The development of a simple method for measuring the HKA angle without radiation exposure will allow quantitative routine evaluation of the


progression of knee OA and its postoperative course. Conventional motion analysis has mainly used inertial motion capture systems and optical marker motion capture systems. Marker motion


capture requires optical reflective markers and has high measurement accuracy, but also requires a lot of space, time and cost. Recently, markerless motion capture methods, such as methods


to create three-dimensional (3D) images from multiple camera images17, methods using depth sensors18,19 and methods using deep learning20,21, have been validated for accuracy and clinical


application. OpenPose, an algorithm for estimating two-dimensional (2D) human poses in images using deep learning, is one such method22,23. OpenPose can estimate body feature points from


photographs and videos using a standard digital camera without special measurement equipment or environment. This open-source software is available for free non-commercial use and is


increasingly being applied in the field of motion analysis because of its ease of implementation20,24,25. These reports on healthy participants have suggested that OpenPose could be a useful


tool for measuring knee joint range of motion in the sagittal plane. However, the measurement accuracy of the HKA angle in the coronal plane using OpenPose has not been validated.


Furthermore, it is important in terms of clinical application to validate the accuracy of OpenPose for measuring HKA angle in deformed lower limbs, such as osteoarthritic knees, not just


healthy knees. Verification of its reliability and validity may lead to the establishment of a new, simple, and inexpensive measurement tool in clinical practice and at home. This


verification also has the potential to make it easier than ever to investigate changes in knee alignment over time. Therefore, this prospective study aimed to clarify the reliability and


validity of HKA angle measurements using OpenPose pre- or post-operative TKA compared with radiography. We hypothesised that OpenPose would be a highly reliable and valid HKA angle


measurement tool. METHODS STUDY DESIGN AND PATIENTS In this prospective study, we included patients who underwent TKA for knee OA between April 2021 and February 2022. The sole exclusion


criterion was inability to maintain a standing position because this might interfere with accurate measurements. However, none of the cases met this exclusion criterion. A total of 60 knees


(30 patients) were assessed with pre- or post-operative radiographs. Out of the 60 knees with OA, 35 were TKA knees and 25 were non-TKA knees. Patients were assessed either pre- or post-


operatively, not both. All procedures were performed using a cruciate-retaining Persona knee system (Zimmer Biomet Inc., Warsaw, IN) by a single surgeon at a single institution. This study


was approved by Nittaduka Medical and Welfare Center Ethics Review Committee and performed according to the principles of the Declaration of Helsinki. All patients provided written informed


consent. We used GPower 3.1.9.2 to determine the appropriate sample size. In the linear regression analysis, a sample size of 55 participants was required to obtain an effect size (f2)


greater than 0.15 with 0.80 power and 0.05 error level on two-tailed test. Therefore, this study enrolled 60 knees. Participants were asked to wear plain trousers of appropriate size to


prevent misinterpretation of OpenPose measurements. DATA COLLECTION AND ANALYSIS We collected two types of imaging data for HKA angle estimation, first with radiography then with OpenPose,


pre- or post-operatively. We obtained AP full-length radiographs with the patient in a standing position with the knee extended and the patella facing forward. The X-ray beam was


perpendicular to the plane of the leg and at the level of the knee joint. Next, we obtained coronal images of the lower extremity using a digital camera immediately after the radiography.


The patient’s posture and imaging direction were the same as for the radiography; the images were taken perpendicular to the plane of the leg, at the level of the knee joint, so that the


entire body was captured on screen. The two types of images were then analysed. For radiography, the HKA angles were measured twice from the same radiographs with a 1-week interval between


measurements. The average value of the two measurements was calculated. The HKA angle was defined as the angle between the femoral and tibial mechanical axis (Fig. 1A)1. The femoral


mechanical axis was defined as a line connecting the centre of the femoral head and knee joint. The tibial mechanical axis was defined as a line connecting the centre of the knee and ankle


joint. The HKA angles were expressed as degrees deviation from 180° and characterised as being varus (≤ − 3°), neutral (0 ± 3°), or valgus (≥ 3°)3,4. This measurement was conducted by two


separate raters. To prevent information bias, the second rater was blinded to the OpenPose data. The data of the second rater were used as the representative values. The inter-rater


reliability between the two raters was evaluated and found to be excellent (intraclass correlation coefficient [ICC] = 0.994); thus, the radiographic measurement by the second rater was


deemed to be sufficiently reliable. For OpenPose, the feature points of each joint were estimated from the relevant images (Fig. 1B). The HKA angles were calculated from the estimated


coordinates of the hip, knee, and ankle joints. These analyses were performed twice from the same image automatically by Python code, and the average values were compared with those of the


radiographic measurements. STATISTICAL ANALYSIS The test–retest reliability was evaluated using ICCs (1, 1). ICC values of < 0.5, 0.5–0.75, 0.75–0.9, and > 0.9 indicated poor,


moderate, good, and excellent agreement, respectively26. The OpenPose measurement was compared with radiographic measurement to assess its validity. Linear regression analysis was used to


evaluate the possibility of estimating the radiographic value from the OpenPose value. R2 values of 0.01–0.09, 0.09–0.25, and > 0.25 indicated small, medium, and large correlations,


respectively20. ICCs (2, 1) and Bland–Altman analysis were used to evaluate the agreement between the OpenPose and radiographic values27,28. The mean of the difference between the


radiographic and OpenPose measurements is a fixed error, and the mean of the absolute difference between the radiographic and OpenPose measurements is an absolute error. All statistical


analyses were performed using R software version 4.1.0 with a significance level of _p_ < 0.0529. RESULTS We analysed 60 knees (22 men and 38 women; mean age: 73.8 ± 6.0 years; mean


height: 156.8 ± 8.0 cm; mean mass: 64.7 ± 8.8 kg; body mass index: 26.3 ± 3.0 kg/m2; TKA knees: 35, OA knees: 25; varus knees: 18, neutral knees: 39, valgus knees: 3). The mean (standard


deviation [SD]) HKA angles were − 1.59° (5.67) and − 2.67° (5.90) for OpenPose and radiography, respectively. The mean (SD) of the difference between OpenPose and radiography is − 1.077°


(2.171). The mean (SD) of the absolute difference between OpenPose and radiography is 1.799° (1.613). The test–retest reliability values (ICCs) of these two measurements are listed in Table 


1. Both measurements showed excellent test–retest reliability. The estimation possibility and agreement between the measurements are shown in Table 2. The regression analysis indicated a


large correlation. The ICC (2, 1) between OpenPose and radiography indicated excellent agreement (0.915). Results of the Bland–Altman analysis and the Bland–Altman plot are shown in Table 3


and Fig. 2, respectively. A significant fixed error (mean difference [MD]) of − 1.077° was observed between OpenPose and radiography. No significant proportional error was observed between


OpenPose and radiography. The random error (SD for MD) of OpenPose against radiography was 2.171. DISCUSSION This study investigated the clinical efficacy, reliability and validity of HKA


angle measurement using OpenPose. The main findings of this study were that OpenPose had excellent test–retest reliability (ICC (1, 1) = 1.000) and excellent agreement with radiography (ICC


(2, 1) = 0.915) for HKA angle measurement. The random error of OpenPose against radiography was small at 2.171. This study was the first to apply OpenPose to HKA angle measurement. OpenPose


uses models trained on a large number of images to predict a confidence map for each joint at each pixel in a 2D image22. OpenPose takes the maximum value of the confidence map to


distinguish the accuracy of peaks and considers the pixel with the maximum value to be the joint position. This mechanical estimation of joint positions may allow for strongly reproducible


measurements from the same image. OpenPose measurement had large correlation and excellent agreement with radiography, thus confirming its validity. OpenPose had a fixed error of 1.077°


underestimation compared to the gold standard full-length radiography, an absolute error of 1.799°, and a random error of 2.171°. Previous studies comparing the measurement of HKA angle


using full-length radiographs vs standardized AP knee radiographs reported a fixed error of 1.8–4.4°, an absolute error of 2.5°, and a random error of 3.6°15,16,30,31. The misalignment of


the mechanical and anatomical axes of the femur and tibia are thought to have produced these errors between standard AP and full-length radiographic measurements. Additionally, Vanwanseele


et al.32 reported that HKA angle measurement by the marker-based 3D motion analysis system had a fixed error of 3.6° underestimation when compared with full-length radiographic measurement


and a random error of 2.4°. This fixed error may be due to the method used to estimate the hip centre. The method used by Vanwanseele tends to estimate the hip centre externally, which could


have resulted in a larger fixation error in the valgus direction33. Furthermore, the random error of the marker-based 3D motion analysis system was comparable to that of OpenPose in the


present study. Therefore, in terms of validity and reliability, OpenPose has the potential to provide just as good if not better HKA angle measurements than other systems under controlled


measurement tasks and conditions; OpenPose is close to the calculated values of the gold standard radiographic measurements for HKA angle measurements. OpenPose also underestimated HKA angle


by 1.077° in the varus direction, which may be because it estimates joint position based on differences in the contrast of pixels around the joint. The hip joint, which has more soft


tissue, may have larger estimation errors than the knee and ankle joints. However, the effect of underestimation can be reduced by offsetting − 1.077° from the output value. The mechanical


axis after TKA is ideally within 0° ± 3° in terms of loosening34. Although an error of − 1.077° is not large, it has a SD of 2.171°. Therefore, without the offset of − 1.077° from the output


value, more cases are likely to be incorrectly determined as having malalignment than with the offset. We thus encourage users to offset the output to improve the validity of OpenPose. The


accuracy of OpenPose estimation in this study was higher than in previous reports20,24,35. Ota et al. reported an ICC (2, 1) of 0.8 for maximum knee flexion angle in squatting and 0.62–0.67


for maximum knee flexion angle in walking on a treadmill with regard to the measurement validity of OpenPose compared with a 3D motion analysis system20,24. Yamamoto et al.35 similarly


reported an ICC (2, 1) of 0.54–0.67 for maximum knee flexion angle during walking on a treadmill. Squatting differs from walking in that the positional relationship between the subject and


the camera is less likely to change. The plane formed by the three joint points must be aligned with the plane to be measured to accurately measure joint angles with a single camera.


Therefore, 2D images must be taken from a direction perpendicular to the plane formed by the three joint points to measure the joint angles accurately with a single camera. The image for


OpenPose was taken perpendicular to the plane of the leg at the height of the knee joint, as in radiography. As a result, the OpenPose measurement may have had a high estimation accuracy.


Therefore, if the task conditions (i.e., camera and subject positioning) are easy to control, OpenPose has the ability to produce highly accurate measurements, which increases its potential


for clinical and research applications. OpenPose measurements do not provide information regarding whether a lower extremity malalignment is due to intra- or extra-articular deformities;


therefore, its use is not recommended when the aim is to evaluate intra-articular or extra-articular deformities. However, OpenPose allows for frequent measurements without radiation


exposure and can play a valuable role in assessing the progression of knee OA and its post-operative course. In the future, OpenPose has the potential to be a cost-effective portable


measurement tool suitable for the frequent evaluation of lower extremity alignment in clinical and home settings. This study has several limitations. First, the sample size of valgus knee


deformities was small. However, OpenPose estimates the knee joint position based on differences in pixel contrast around the knee joint. Therefore, whether the knee joint is positioned


medially or laterally relative to the hip or ankle joint may not affect the estimation accuracy of varus or valgus deformities. In the future, additional validation should be carried for


valgus deformities. Second, it was impossible to perform the radiographic and OpenPose imaging simultaneously. Nonetheless, we instructed patients not to move until the two measurements were


completed, and the two measurements were performed consecutively. Furthermore, we ensured that the lower limbs did not move until these measurements were completed. Third, the two


measurements were not performed in a randomised order. However, HKA angle measurements from the radiographic images were estimated by a second rater who was blinded to the OpenPose


measurements. OpenPose measurement was also performed mechanically. Therefore, we believe that information bias was unlikely to have been introduced by the order of measurement. Fourth, the


accuracy of OpenPose estimation may be affected by the background. To validation the impact that different backgrounds may have had on the results of the OpenPose measurement, we measured by


transposing different backgrounds onto the images of patients. We found that there was good agreement in the results and minimal differences in the errors. Therefore, it is unlikely that


the fact that the background of this experiment is a radiology room will diminish the clinical applicability of OpenPose. In conclusion, this study supported our hypothesis that OpenPose is


a reliable and valid tool for measuring HKA angles in patients with knee OA. OpenPose, which enables non-invasive and simple measurements, may be a useful tool to assess changes in HKA angle


and monitor the progression and post-operative course of knee OA. Furthermore, this validated tool can be used not only in clinics and hospitals, but also at home and in training gyms, and


thus has the potential for widespread application. DATA AVAILABILITY The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable


request. REFERENCES * Cooke, D. _et al._ Axial lower-limb alignment: comparison of knee geometry in normal volunteers and osteoarthritis patients. _Osteoarthr. Cartil._ 5, 39–47 (1997).


Article  CAS  Google Scholar  * Tarassoli, P. _et al._ Arithmetic hip-knee-ankle angle and stressed hip-knee-ankle angle: Equivalent methods for estimating constitutional lower limb


alignment in kinematically aligned total knee arthroplasty. _Knee Surg Sports Traumatol Arthrosc_ 30, 2980–2990 (2022). Article  PubMed  PubMed Central  Google Scholar  * Shelton, T. J.,


Nedopil, A. J., Howell, S. M. & Hull, M. L. Do varus or valgus outliers have higher forces in the medial or lateral compartments than those which are in-range after a kinematically


aligned total knee arthroplasty? Limb and joint line alignment after kinematically aligned total knee arthroplasty. _Bone Jt. J._ 99, 1319–1328 (2017). Article  Google Scholar  * Bellemans,


J., Colyn, W., Vandenneucker, H. & Victor, J. The Chitranjan Ranawat Award: Is neutral mechanical alignment normal for all patients? The concept of constitutional varus. _Clin. Orthop.


Relat. Res._ 470, 45–53 (2012). Article  PubMed  Google Scholar  * Sharma, L. _et al._ The role of knee alignment in disease progression and functional decline in knee osteoarthritis. _JAMA_


286, 188–195 (2001). Article  CAS  PubMed  Google Scholar  * Goto, N. _et al._ Alignment factors affecting the medial meniscus extrusion increases the risk of osteoarthritis development.


_Knee Surg. Sports Traumatol. Arthrosc._ 27, 2617–2623 (2019). Article  PubMed  Google Scholar  * Tanamas, S. _et al._ Does knee malalignment increase the risk of development and progression


of knee osteoarthritis? A systematic review. _Arthritis Rheumatol._ 61, 459–467 (2009). Article  Google Scholar  * Hayashi, D. _et al._ Knee malalignment is associated with an increased


risk for incident and enlarging bone marrow lesions in the more loaded compartments: The MOST study. _Osteoarthr. Cartil._ 20, 1227–1233 (2012). Article  CAS  Google Scholar  * Brouwer, G.


M. _et al._ Association between valgus and varus alignment and the development and progression of radiographic osteoarthritis of the knee. _Arthritis Rheum_ 56, 1204–1211 (2007). Article 


CAS  PubMed  Google Scholar  * Matsuda, S., Kawahara, S., Okazaki, K., Tashiro, Y. & Iwamoto, Y. Postoperative alignment and ROM affect patient satisfaction after TKA. _Clin. Orthop.


Relat. Res._ 471, 127–133 (2013). Article  PubMed  Google Scholar  * Howell, S. M., Shelton, T. J. & Hull, M. L. Implant survival and function ten years after kinematically aligned total


knee arthroplasty. _J. Arthroplasty_ 33, 3678–3684 (2018). Article  PubMed  Google Scholar  * Lee, B. S., Cho, H. I., Bin, S. I., Kim, J. M. & Jo, B. K. Femoral component varus


malposition is associated with tibial aseptic loosening after TKA. _Clin. Orthop. Relat. Res._ 476, 400–407 (2018). Article  PubMed  PubMed Central  Google Scholar  * Liu, H. X., Shang, P.,


Ying, X. Z. & Zhang, Y. Shorter survival rate in varus-aligned knees after total knee arthroplasty. _Knee Surg. Sports Traumatol. Arthrosc._ 24, 2663–2671 (2016). Article  PubMed  Google


Scholar  * Ritter, M. A. _et al._ The effect of alignment and BMI on failure of total knee replacement. _J. Bone Jt. Surg. Am._ 93, 1588–1596 (2011). Article  Google Scholar  * Borton, Z.,


Shivji, F., Eyre-Brook, A. I., Wilson, A. & Yasen, S. Non-weightbearing imaging and standard knee radiographs are inferior to formal alignment radiographs for calculating coronal


alignment of the knee. _Radiography_ 27, 260–265 (2021). Article  CAS  PubMed  Google Scholar  * Abu-Rajab, R. B. _et al._ Hip–knee–ankle radiographs are more appropriate for assessment of


post-operative mechanical alignment of total knee arthroplasties than standard AP knee radiographs. _J. Arthroplasty_ 30, 695–700 (2015). Article  PubMed  Google Scholar  * Hurley, R. J.,


Davey, M. S., Newell, M. & Devitt, A. Assessing the accuracy of measuring leg length discrepancy and genu varum/valgum using a markerless motion analysis system. _J. Orthop._ 26, 45–48


(2021). Article  PubMed  PubMed Central  Google Scholar  * Yeung, L. F., Yang, Z., Cheng, K. C. C., Du, D. & Tong, R. K. Y. Effects of camera viewing angles on tracking kinematic gait


patterns using Azure Kinect, Kinect v2 and Orbbec Astra Pro v2. _Gait Posture_ 87, 19–26 (2021). Article  PubMed  Google Scholar  * Thomas, J., Hall, J. B., Bliss, R. & Guess, T. M.


Comparison of Azure Kinect and optical retroreflective motion capture for kinematic and spatiotemporal evaluation of the sit-to-stand test. _Gait Posture_ 94, 153–159 (2022). Article  PubMed


  Google Scholar  * Ota, M., Tateuchi, H., Hashiguchi, T. & Ichihashi, N. Verification of validity of gait analysis systems during treadmill walking and running using human pose tracking


algorithm. _Gait Posture_ 85, 290–297 (2021). Article  PubMed  Google Scholar  * Iijima, H. _et al._ Contralateral limb effect on gait asymmetry and ipsilateral pain in a patient with knee


osteoarthritis: A proof-of-concept case report. _JBJS Case Connect_ 10, e0418 (2020). Article  PubMed  Google Scholar  * Cao, Z., Simon, T., Wei, SE., Sheikh, Y. Realtime multi-person 2D


pose estimation using part affinity fields. In _Proceedings of the IEEE Conference on Computer Vision and Pattern _Recognition, 7291–7299 (2017). * Cao, Z., Hidalgo, G., Simon, T., Wei, S.


E. & Sheikh, Y. OpenPose: real time multi-person 2D pose estimation using Part Affinity Fields. _IEEE Trans. Pattern Anal. Mach. Intell._ 43, 172–186 (2019). Article  Google Scholar  *


Ota, M. _et al._ Verification of reliability and validity of motion analysis systems during bilateral squat using human pose tracking algorithm. _Gait Posture_ 80, 62–67 (2020). Article 


PubMed  Google Scholar  * Stenum, J., Rossi, C. & Roemmich, R. T. Two-dimensional video-based analysis of human gait using pose estimation. _PLOS Comput. Biol._ 17, e1008935 (2021).


Article  ADS  CAS  PubMed  PubMed Central  Google Scholar  * Koo, T. K. & Li, M. Y. A guideline of selecting and reporting intraclass correlation coefficients for reliability research.


_J. Chiropr. Med._ 15, 155–163 (2016). Article  PubMed  PubMed Central  Google Scholar  * Bland, J. M. & Altman, D. G. Statistical methods for assessing agreement between two methods of


clinical measurement. _Lancet_ 1, 1307–1310 (1986). Google Scholar  * Bland, J. M. & Altman, D. G. Measuring agreement in method comparison studies. _Stat. Methods Med. Res._ 8, 135–160


(1999). Article  CAS  PubMed  Google Scholar  * R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing, Vienna, Austria.


https://www.r-project.org/ Cited 18 October 2022. * Graden, N. R., Dean, R. S., Kahat, D. H., DePhillipo, N. N. & LaPrade, R. F. True mechanical alignment is found only on full-limb and


not on standard anteroposterior radiographs. _Arthrosc. Sports Med. Rehabil._ 2, e753–e759 (2020). Article  PubMed  PubMed Central  Google Scholar  * Van Raaij, T. M., Brouwer, R. W.,


Reijman, M., Bierma-Zeinstra, S. M. & Verhaar, J. A. Conventional knee films hamper accurate knee alignment determination in patients with varus osteoarthritis of the knee. _Knee_ 16,


109–111 (2009). Article  PubMed  Google Scholar  * Vanwanseele, B., Parker, D. & Coolican, M. Frontal knee alignment: three-dimensional marker positions and clinical assessment. _Clin.


Orthop. Relat. Res._ 467, 504–509 (2009). Article  PubMed  Google Scholar  * Bell, A. L., Pedersen, D. R. & Brand, R. A. A comparison of the accuracy of several hip center location


prediction methods. _J. Biomech._ 23, 617–621 (1990). Article  CAS  PubMed  Google Scholar  * Jeffery, R. S., Morris, R. W. & Denham, R. A. Coronal alignment after total knee


replacement. _J. Bone Jt. Surg._ 73, 709–714 (1991). Article  CAS  Google Scholar  * Yamamoto, M. _et al._ Accuracy of temporo-spatial and lower limb joint kinematics parameters using


OpenPose for various gait patterns with orthosis. _IEEE Trans. Neural Syst. Rehabil. Eng._ 29, 2666–2675 (2021). Article  PubMed  Google Scholar  Download references ACKNOWLEDGEMENTS We wish


to thank the radiological technologists in our hospital for their help with the radiographic assessment. AUTHOR INFORMATION AUTHORS AND AFFILIATIONS * Department of Orthopaedic Surgery,


Graduate School of Medical Sciences, Kanazawa University, 13-1 Takaramachi, Kanazawa City, Ishikawa, 920-8641, Japan Yoshitomo Saiki, Tamon Kabata, Yoshitomo Kajino, Daisuke Inoue, Takaaki


Ohmori, Junya Yoshitani, Takuro Ueno, Yuki Yamamuro, Atsushi Taninaka, Tomoyuki Kataoka & Hiroyuki Tsuchiya * Department of Rehabilitation Physical Therapy, Faculty of Health Science,


Fukui Health Science University, 55-13-1 Egami, Fukui City, Fukui, 910-3190, Japan Yoshitomo Saiki * Department of Orthopaedic Surgery, Fukui General Hospital, 58-16-1 Egami, Fukui City,


Fukui, 910-8561, Japan Tomohiro Ojima & Seigaku Hayashi * Department of Rehabilitation Physical Therapy, Fukui General Hospital, 58-16-1 Egami, Fukui City, Fukui, 910-8561, Japan Naoyuki


Kubo Authors * Yoshitomo Saiki View author publications You can also search for this author inPubMed Google Scholar * Tamon Kabata View author publications You can also search for this


author inPubMed Google Scholar * Tomohiro Ojima View author publications You can also search for this author inPubMed Google Scholar * Yoshitomo Kajino View author publications You can also


search for this author inPubMed Google Scholar * Daisuke Inoue View author publications You can also search for this author inPubMed Google Scholar * Takaaki Ohmori View author publications


You can also search for this author inPubMed Google Scholar * Junya Yoshitani View author publications You can also search for this author inPubMed Google Scholar * Takuro Ueno View author


publications You can also search for this author inPubMed Google Scholar * Yuki Yamamuro View author publications You can also search for this author inPubMed Google Scholar * Atsushi


Taninaka View author publications You can also search for this author inPubMed Google Scholar * Tomoyuki Kataoka View author publications You can also search for this author inPubMed Google


Scholar * Naoyuki Kubo View author publications You can also search for this author inPubMed Google Scholar * Seigaku Hayashi View author publications You can also search for this author


inPubMed Google Scholar * Hiroyuki Tsuchiya View author publications You can also search for this author inPubMed Google Scholar CONTRIBUTIONS Y.S.: conceptualisation, methodology,


investigation, data curation, formal analysis, writing—original draft. T.K.: conceptualisation, methodology, supervision, writing—review and editing. T.O.: conceptualisation, methodology,


supervision, writing—review and editing. Y.K.: methodology, investigation, supervision. D.I.: methodology, investigation, supervision. T.O.: methodology, supervision. J.Y.: methodology,


supervision. T.U.: methodology, supervision. Y.Y.: methodology, supervision. A.T.: methodology, supervision. T.K.: methodology, supervision. N.K.: investigation, validation. S.H.:


supervision, writing—review and editing. H.T.: conceptualisation, supervision, writing—review and editing. CORRESPONDING AUTHOR Correspondence to Tamon Kabata. ETHICS DECLARATIONS COMPETING


INTERESTS The authors declare no competing interests. ADDITIONAL INFORMATION PUBLISHER'S NOTE Springer Nature remains neutral with regard to jurisdictional claims in published maps and


institutional affiliations. RIGHTS AND PERMISSIONS OPEN ACCESS This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing,


adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons


licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise


in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the


permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. Reprints and


permissions ABOUT THIS ARTICLE CITE THIS ARTICLE Saiki, Y., Kabata, T., Ojima, T. _et al._ Reliability and validity of OpenPose for measuring hip-knee-ankle angle in patients with knee


osteoarthritis. _Sci Rep_ 13, 3297 (2023). https://doi.org/10.1038/s41598-023-30352-1 Download citation * Received: 09 December 2022 * Accepted: 21 February 2023 * Published: 25 February


2023 * DOI: https://doi.org/10.1038/s41598-023-30352-1 SHARE THIS ARTICLE Anyone you share the following link with will be able to read this content: Get shareable link Sorry, a shareable


link is not currently available for this article. Copy to clipboard Provided by the Springer Nature SharedIt content-sharing initiative


Trending News

Buena Park : Early Morning Mall Fire Destroys Liquor Store

A fire broke out early Saturday in a neighborhood mini-mall, destroying a liquor store and seriously damaging two adjace...

Tom Hayden Travels Back Into His Past

To be sure, there was the Free Speech movement of Berkeley, and love power, and the misguided promise of freedom through...

Eu funds wrangling threatens regions - farmers weekly

23 February 1998 EU FUNDS WRANGLING THREATENS REGIONS BRITAINS regions face the prospect of a serious delay in receiving...

'polished, assertive, on the attack': what to know about karen read's trial style

Crime “THERE’S AN OVERALL PERSONA THAT SHE'S VERY MUCH PUSHING THAT IS NOT WHAT ONE WOULD NORMALLY DO IN A CASE LIK...

Figures show a better future for uk pigs - farmers weekly

_BY PETER CRICHTON_ ACCORDING to recent statistics, the UK pig industry faces a better future than a year ago. After a p...

Latests News

Reliability and validity of openpose for measuring hip-knee-ankle angle in patients with knee osteoarthritis

ABSTRACT We aimed to assess the reliability and validity of OpenPose, a posture estimation algorithm, for measuring hip-...

Retrosense therapeutics welcomes dr. Steven bramer as chief development officer

ANN ARBOR, Mich.--(BUSINESS WIRE)-- RetroSense Therapeutics, a biotechnology company dedicated to developing gene therap...

Class 3 medicines recall: intrapharm laboratories ltd, kolanticon gel 500ml, el (21)a/01

CLASS 3 MEDICINES RECALL: INTRAPHARM LABORATORIES LTD, KOLANTICON GEL 500ML, EL (21)A/01 Intrapharm Laboratories Ltd has...

Can a world series win help boston's healing?

RACHEL MARTIN, HOST: Time now for sports. (SOUNDBITE OF MUSIC) MARTIN: Well, if you didn't see it live, you've...

50,000 stray dogs in detroit? That 'makes no sense'

A disturbing story from Bloomberg News that we posted about on Wednesday — " Detroit's Stray Dog Epidemic: 50,...

Top